QA of CT Scanner using water phantom
A
good Quality Assurance (QA) program will provide regular testing, prompt
interpretation of test data, and faithful record keeping. A current logbook or
computer file should be readily available for viewing by any regulatory agency
in the event of an unscheduled inspection. In addition to the following tests,
daily visual checks should be done on table/gantry movement, cables, cords,
operating console, controls, and print system if installed. These visual
tests should be documented as well, for future comparison.
Facilities with equipment under warranty or
service contract with an Original Equipment Manufacturer (OEM) or an
Independent Service Organization (ISO) must follow the testing and preventive
maintenance schedule required by the OEM or ISO to keep the warranty or
contract valid. The OEM or ISO testing and maintenance schedule must be
included in the manual. Facilities with equipment not under warranty or service
contract must follow the testing frequency stated in this guide. Facilities
must perform all the QC tests which their manufacturer supplied phantom will
allow.
Appropriate
quality control testing must be conducted whenever major maintenance (X-ray
tube replacement) or a change in equipment operation (software change) occurs.
1. Testing frequency - Each day of
operation
Equipment functioning: Each day during
equipment warm-up, and before scanning the first patient, the operator must
check for any malfunction. The operator must also evaluate the mechanical and
electrical safety of the CT system. Malfunctions and unsafe conditions shall be
corrected promptly.
CT Number Accuracy of Water, Image Noise,
Image Uniformity, and Artifact evaluation shall be performed daily using a
common technique setting with a head sized phantom. The average CT number and
standard deviation of noise must be performed using a region of interest that
is standardized for the QA tests. Image uniformity and artifact evaluation must
be done using an appropriate and consistent window level and width.
This is the water phantom used in daily
QA carried out by radiographer to test the CT number accuracy of water, image
noise, image uniformity and artifact.
CT Number
for Water
|
This chart shows
the various CT numbers (Hounsfield units) calculated for various tissues and
substances based on the density of water. Notice that at the extremes is bone
(+1000 HU) and air (-1000 HU). Water has a CT number of zero, which is used to
test for the function of the algorithm that calculates CT numbers.
2. Frequency - Monthly
Imaged Slice Thickness or Slice
Sensitivity Profile
Performed at each available slice
thickness, using an appropriate phantom for both the axial and helical modes of
operation.
Slice Positioning Accuracy
Evaluate the accuracy of slice
localization lights, the accuracy of slice positioning and the accuracy of table
motion with incrementation.
CT Number Scale Accuracy
CT numbers must be evaluated using a
phantom containing a variety of materials to evaluate a wide range of CT
numbers. The measured values must remain within manufacturer's specifications.
Hard Copy Devices
Hard copy devices and dry image processors
must be monitored for consistency in accordance with the manufacturers
published recommendations. Many of these devices utilize a SMPTE test pattern,
which can be readily used and printed for review.
3. Test frequency - Semi annually
Dose Profile Width
This test must be performed at each
available slice thickness. This test can readily be performed using packaged
film, which is placed around the surface of the phantom. Alternative methods of
evaluation can also be used. Multiple slice units need to be evaluated for
over-beaming and adjusted to manufacturer's specifications.
Spatial Resolution
Must be evaluated in the x-y plane using a
phantom with suitable objects. The objects will be of different sizes and
shapes. In-plane resolution must be evaluated in the axial mode, for a single
slice thickness using both standard and high-resolution algorithms. If an
appropriate phantom is available, the system must also be evaluated in the helical
mode to test for longitudinal resolution with different combinations of
collimation and pitch.
Low-Contrast Detectability
Must be performed using a phantom with
test objects that will evaluate the system for less than 1% contrast
sensitivity.
4. Test frequency - Annually
Patient dose
Each facility shall have available dose
measurements based on the most common conditions of operation of their CT
units. Since the advent of CTDI FDA, the International
Electrotechnical Commission (IEC) has defined a new standard for dose index.
This standard is called the CTDI100. CTDI100 utilizes measurements made
with a 16cm diameter (head/pediatric body) or a 32cm diameter (body) acrylic
phantom. The measurements are made utilizing a 100mm long pencil ionization
chamber. Readings are made with the ion chamber in both the center (axial or
central dose) position and near surface slots of the phantom (the peripheral
dose).
CTDI W, the weighted or blended
dose, is calculated by adding together two-thirds of the CTDI 100 peripheral
dose with one-third of the CTDI 100 axial or center dose.
(CTDI W = 2/3 CTDI 100 peripheral + 1/3 CTDI 100
axial or center
It is important to remember that
CTDI W represents an average dose in the x and y planes. With
multiple slice helical scanning, the dose is better represented by
CTDIVOL which takes into account the effect of pitch and averages over the
x, y, and z planes. CTDI VOL represents the integrated dose over the
total volume that is irradiated.
CTDI VOL = 1 ¸ PITCH X
CTDI W
The resultant measurements should not
exceed the following:
Examination
|
CTDIVOL (mGy)
|
Adult Head
|
75
|
Adult Abdomen
|
25
|
Pediatric Abdomen (5 yr. Old)
|
20
|
And the measurements shall not exceed:
Examination
|
CTDIVOL (mGy)
|
Adult Head
|
80
|
Adult Abdomen
|
30
|
Pediatric Abdomen (5 yr. Old)
|
25
|
5. On Installation of New Tube
All daily and monthly tests must be
completed before patient examinations commence. All semi-annual and annual
tests must be completed within thirty days of tube replacement.
6. On Installation of New Unit
All daily and monthly tests must be
completed before patient examinations commence. All semi-annual and annual
tests must be completed within thirty days of a new installation.
Half-Value Layer
The manufacturer's procedure to evaluate
HVL shall be followed and the manufacturer's published specifications shall be
met.
7. Radiation Protection Survey
A radiation protection survey must be
completed before operation to ensure that the structural shielding is adequate
to meet the safety requirements.
Scan protocols
Each CT unit shall have posted scan
protocols available for reference by the operator. As a minimum, these
protocols shall be include the following scan factors: use of oral contrast;
use of intravenous contrast; slice thickness; slice spacing; anatomical
description of the extent of the field for scanogram; and the technical factors
selected for the examination. Technical factors must differentiate between
pediatric and adult sized patients as well as anatomical areas that are to be
imaged.
The radiologist, CT technologist and
medical physicist should work together on establishing the technique (kV and
mAs) used on each projection, so that patient dose is minimized. The technique
used should always be sufficient, so as to avoid the appearance of clinically
unacceptable quantum mottle.
Nice post
ReplyDeleteEasybookmylab.com helps you to find the CT scan centre in Mumbai among the many scan centres available nearby your location at best price.