Conventional vs CT-SIM





Differences between Conventional and CT simulation

Differences
Conventional simulation
Computed tomography simulation
Treatment position
Identified with to bony landmarks under fluoroscopic condition
Identified to bony landmarks through the use of scout+pilot image
Identification of target volume
Identified from bony landmarks
Can get directly from CT data
Determination of beam geometry
-Verified using fluoroscopic
-Doesn’t have standard beam geometry
-Verified from BEV/DRR
-Have standard beam geometry
Contour acquisition
Patient contour is obtained through the use of lead wire (manual)
Patient contour is obtained from CT data
Shielding design
-From bony landmarks
-Cannot use with beam geometry
-Conformal to target
-Can be used together with the beam geometry
Virtual simulation
No
Yes
Image of radiograph
Have limited soft tissue contrast
Have good soft tissue contrast







Machine for Conventional and CT simulator



A conventional treatment simulator



A dedicated radiotherapy CT simulator



SUMMARY OF THE CONVENTIONAL SIMULATION
PROCEDURE FOR A TYPICAL PATIENT (SIX STEPS)

step
Conventional simulation
1
Determination of patient treatment position with fluoroscopy
2
Determination of beam geometry
3
Determination of field limits and isocentre
4
Acquisition of contour
5
Acquisition of BEV and set-up radiographs
6
Marking of patient





SUMMARY OF THE PROCEDURE FOR A TYPICAL
PATIENT COMPUTED TOMOGRAPHY SIMULATION (NINE STEPS)

step
CT simulation
1
Determination of patient treatment position with pilot/scout films
2
Determination and marking of reference isocentre
3

Acquisition of CT data and transfer to virtual simulation workstation
4
Localization and contouring of targets and critical structures
5
Determination of treatment isocentre with respect to target and reference
isocentre
6
Determination of beam geometry
7
Determination of field limits and shielding
8
Transfer of CT and beam data to the TPS
9
Acquisition of BEV and set-up DRRs






Images



Pilot or scoutimages relate slice position to radiographic landmarks





DRR. Note that grey levels, brightness and contrast can be adjusted to
provide an optimal image





A DRR with superimposed BEV for a vertex field of a brain patient. This
treatment geometry would be impossible to simulate on a conventional simulator.



Treatment field in a conventional simulation